The excursion of the tubes in the Coriolis meter is detected by sensors located at the meter inlet and outlet, and is measured as a time delay, or phase shift. The phase shift between the rotational frequencies of both tubes is directly proportional to the mass flow rate.

The resonant frequency of the delta-tubed Coriolis mass flow meter’s tubes changes in proportion to the density of the fluid media. This effect allows for the measurement of media density. Using only one sensor, both density and temperature measurements can be collected, as the extent of deformation of the pipes is temperature-dependent. Therefore, the temperature is measured for compensation purposes. Using a single instrument, values such as mass flow, density, and temperature may be measured.

The ALCM-DT delta-tubed Coriolis mass flow meter’s circuitry also makes it possible to determine conversions such as volumetric flow, percent concentration, and percent water-cut.