Oxygen Analyzer with Remote Sensor GPR-1900 & GPR-2900- Máy phân tích oxy với cảm biến từ xa GPR-1900 & GPR-2900
Nhà cung cấp: Pitesco Việt Nam
Hãng sản xuất: AII
Oxygen Analyzer with Remote Sensor
GPR-1900 & GPR-2900
A panel mounted oxygen analyser with a remote sensor allows flexibility of installation. The GPR-1900 gas analyzer is suitable for trace oxygen measurements as low as 0-1 ppm; the GPR-2900 measures O2 up to 100%. The remote sensor block allows a very easy oxygen sensor replacement for simple and fast maintenance and low cost of ownership. All versions can be locked manually to a single range or auto-range. The user is always aware of the oxygen measurement range used via the range ID output.
Features of Oxygen Analyzer with Remote Sensor
- Measurement ranges from 0-10ppm up to 0-100% O2
- Barometric pressure and temperature compensation
- Accuracy of better than 2% of range
- Two user configurable alarms
- Range ID output
- 12-28 V DC or mains powered options
- XLT sensor options for CO2 backgrounds
Technical Specifications
GPR-1900 | GPR-2900 | |
---|---|---|
Measurement range | 0-10, 0-100, 0-1000 ppm, 0-1%, 0-25% | 0-1%, 0-5%, 0-10%, 0-25%; optional 0-1%, 0-5%, 0-10%, 0-100% |
Accuracy | < 2% of selected range at constant conditions | |
Response time | T90 < 10 seconds | |
Recovery time | 60 seconds in air to < 10 ppm in < 1 hr on N2 purge | NA |
Sensitivity (LDL) | 0.05 ppm | 0.005% |
Linearity | < 0.5% of scale | |
Sensor model | GPR-11-60-4 XLT-11-24-4 for gases containing > 0.5% CO2 |
|
Sensor life 25ºC (77°F) and 1 atm | 24 months in < 1000 ppm O2 6 months in air |
GPR-11-60-4 60 months in air XLT-11-24-4 24 months in air |
Calibration interval | 30 days | |
Inlet pressure | 0.34 2 barg (5-30 psig) with atmospheric vent | |
Flow rate | 0.5 1.0 l/m (1-2 SCFH) | |
Gas connections | 1/8" compression tube fittings | |
Wetted parts | Stainless steel | Delrin; stainless steel optional |
Display | Graphical LCD 12.7 x 7cm (5 x 2.75”); resolution 0.001 ppm GPR-1900 MS2, 0.01 ppm GPR-1999, 0.001% GPR-2900 | |
Enclosure | Painted aluminum 6" x 4" x 4" panel mount | |
Compensation | Barometric pressure and temperature | |
Signal output | 4-20 mA or 1-5V | |
Range ID | 4-20 mA non-isolated or 1-5V Option: reassign relay contacts (eliminates standard alarms). |
|
Alarms | Two user configurable alarms: magnetic coil relays rated 3A at 100 V AC | |
Operating temperature | GPR sensors: 5°C to 45°C (41°F to 113°F) XLT sensors: -10°C to 45°C (14°F to 113°F) |
|
Power | 100-240 V AC or 12-28 V DC |